Redefining instrument reprocessing
Time to care
Rethink everything you know about instrument reprocessing: Careclave revolutionises the workflow in dental practices and clinics. Discover the advantages of our worldwide unique 4-in-1 Solution:
Maximum safety for patients and personnel alike
Do not make any compromises in your hygiene standards: Careclave ensures the reprocessing of up to 200 handpieces and further instruments in a single practice day.
Wellness for your instruments
Protect your precious instruments against wear: Careclave provides the best-possible care for your transmission instruments and chuck systems.
No space in your decontamination room? No problem!
Free up new space in your decontamination room: Careclave requires up to 45 % less space, as it combines a number of devices in one.
And the best thing: You save money
Benefit from our cost-saving 4-in-1 Solution: Careclave reduces maintenance and service costs and the number of instrument sets you need.
Carebox
Do you need transmission or prophylaxis instruments for the next treatment as quickly as possible? Then reprocess up to 8 hollow body instruments in your Carebox.
Carebox & trays
Would you like to enhance the efficiency of your staff? Then benefit from reprocessing and lubrication in the Carebox and sterilisation of up to 8 trays for further instruments in just one cycle.
Long trays
Do you need maximum capacity for the sterilisation of your instruments? Then simply load up to 11 kg of instruments on 8 trays or 4 MELAstore Boxes into Careclave.
Carebox Green
For the safe reprocessing of up to 8 prophylaxis handpieces and ultrasonic scaler tips with torque wrenches.
Carebox Blue
For the safe reprocessing and reliable lubrication of up to 8 turbines and handpieces.

All numbers and facts at a glance
Technical data
With a wide range of different programs, Careclave guarantees optimal instrument reprocessing and lubrication. This section provides information about the Careclave mode (with an inserted Carebox) and Vacuclave mode (without a Carebox):
Careclave mode | Description | Cycle time incl. drying |
---|---|---|
Care-Therm Program | Cleaning inside and outside, thermal disinfection, lubrication (if necessary) of the instruments in Carebox | 16 – 18 min |
Care-S Program | Cleaning inside and outside, lubrication (if necessary), sterilisation (134 °C, holding time 3.5 min) of the instruments in Carebox | 21 – 23 min |
Care-B Combi Program | Cleaning inside and outside, lubrication (if necessary), additional sterilisation (134 °C, holding time 5.5 min) of the instruments in Carebox and on the trays | 33 – 43 min |
Vacuclave mode | Description | Cycle time plus drying* |
---|---|---|
Universal-B Program | Sterilisation (134 °C, holding time 5.5 min) of wrapped or double wrapped instruments up to 11 kg | 13 – 22 min |
Quick-S Program | Sterilisation (134 °C, holding time 3.5 min) of unwrapped instruments up to 7 kg | 9 – 15 min |
Gentle-B Program | Sterilisation (121 °C, holding time 20.5 min) of wrapped or double wrapped instruments up to 7 kg | 27 – 37 min |
Prion-B Program | Sterilisation (134 °C, holding time 20.5 min) of wrapped or double wrapped instruments up to 11 kg | 28 – 37 min |
*DRYtelligence automatically adapts the drying process to your load. Intelligent drying not only achieves the best-possible drying results, but reduces the drying time by up to 80%.
Careclave can be installed quickly and easily in every decontamination room. The following list provides an overview of the installation requirements in terms of electricity, water and compressed air:
Technical data | Careclave 618 |
---|---|
Article number | ME10618 |
Device dimensions (W x H x D) | 48 cm x 50 cm x 65.3 cm, height incl. display 56.2 cm, Careclave fits on a 60 cm deep work surface |
Weight | 69 kg |
Chamber size / volume | Ø 25 cm x 35 cm long / 18 L |
Chamber load quantity | max. 11 kg instruments / 2 kg textiles |
Carebox load quantity | max. 8 hollow body instruments |
Power supply | 220-240 V, 50 Hz |
Power consumption | 3 kW |
Compressed air | 4 – 8 bar (compressed air in accordance with ISO 7494-2) |
Feed water supply | max. 35 °C, 1 – 10 bar, additional water treatment required |
Documentation interfaces | USB, network |
"*" indicates required fields